• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME; DERMATOME HANDPIECE, ELECTRICALLY-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME; DERMATOME HANDPIECE, ELECTRICALLY-POWERED Back to Search Results
Model Number N/A
Device Problems Corroded (1131); Naturally Worn (2988); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/31/2023
Event Type  malfunction  
Event Description
It was reported during preventive maintenance the device had a corroded motor, worn screws, worn reciprocation arm, and switch issue.There is no harm or delay reported.There was no patient involvement.Due diligence is complete and there is no additional event information available.No adverse events were reported as a result of this malfunction.Upon investigation, the motor speed was unstable.
 
Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).G2: belgium.Review of the most recent repair record determined the insulation of the cable was damaged, the reciprocation arm and some screws were worn, the motor speed was unstable and the switch was intermittently functioning.The plug harness assembly, reciprocation arm, screws, motor, and switch were replaced and resolved the reported issue.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME HANDPIECE, ELECTRICALLY-POWERED
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key17843404
MDR Text Key324584876
Report Number0001526350-2023-01240
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number60621554
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2023
Was the Report Sent to FDA? No
Date Manufacturer Received09/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/02/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-