ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME; DERMATOME HANDPIECE, ELECTRICALLY-POWERED
|
Back to Search Results |
|
Model Number N/A |
Device Problems
Corroded (1131); Naturally Worn (2988); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/31/2023 |
Event Type
malfunction
|
Event Description
|
It was reported during preventive maintenance the device had a corroded motor, worn screws, worn reciprocation arm, and switch issue.There is no harm or delay reported.There was no patient involvement.Due diligence is complete and there is no additional event information available.No adverse events were reported as a result of this malfunction.Upon investigation, the motor speed was unstable.
|
|
Manufacturer Narrative
|
This complaint is recorded by zimmer biomet under (b)(4).G2: belgium.Review of the most recent repair record determined the insulation of the cable was damaged, the reciprocation arm and some screws were worn, the motor speed was unstable and the switch was intermittently functioning.The plug harness assembly, reciprocation arm, screws, motor, and switch were replaced and resolved the reported issue.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|