The mayfield skull clamp (a1059) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the returned skull clamp was thoroughly inspected, and no defects were found.All manufacturer specifications regarding functional tests were carried out and completed successfully.The pivot lock opens and closes properly and is not loose.There was no movement/play when closing the lock under pressure.Both the teeth and the threads of the base showed no damage, and the torque screw and ratchet extension were in good condition.As a preventive measure, the piston group was opened, cleaned, and reinstalled, and it worked without errors.The ratchet extension could be locked and unlocked.Finally, the skull clamp underwent a successful functional test in accordance with the manufacturer's instructions.Root cause - evaluation found no device deficiencies that would have contributed to the reported complaint.Probable root cause is improper or suboptimal placement of the skull clamp on the patient.No further investigation is required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|