Catalog Number 03666794001 |
Device Problems
Image Display Error/Artifact (1304); No Device Output (1435)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/05/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The test strips' lot number was 71519013 with an expiration date of 31-dec-2024.During the troubleshooting, a display check was performed and revealed the following: 888 appeared clear and complete.The hourglass was faded.The meter memory result appeared faded on the sides.The investigation is ongoing.
|
|
Event Description
|
We received an allegation about a coaguchek xs meter having a faded display and not detecting the test strips.The reporter stated that they had to tilt the meter in order to see everything on the display because it was faded.The reporter also confirmed that no harm occurred in relation to the display issues because the meter had not been in use since the issues began.
|
|
Manufacturer Narrative
|
The meter was not received for investigation.
|
|