• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number R5C4479
Device Problems Fluid/Blood Leak (1250); Material Puncture/Hole (1504); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2023
Event Type  malfunction  
Manufacturer Narrative
E1: initial reporter postal code: (b)(6).A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a homechoice device experienced a system error 2367 (air in line/set) alarm.The patient was connected at the time of the alarm.This occurred during initial drain of peritoneal dialysis therapy.During troubleshooting, it was reported that there was a hole in the homechoice cassette which resulted in a leak that led to this alarm.The leak was further described as ¿liquid coming out the door¿ and ¿fluid all over the floor." renal therapy services (rts) assisted the patient in removing the cassette and advised to start over with new supplies.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park d
singapore 73875 0
SN   738750
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17855242
MDR Text Key324737544
Report Number1416980-2023-04966
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00085412030531
UDI-Public(01)00085412030531
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberR5C4479
Device Lot NumberS23E04019
Was Device Available for Evaluation? No
Date Manufacturer Received09/06/2023
Date Device Manufactured05/04/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
HOMECHOICE CLARIA DEVICE; UNSPECIFIED SOLUTION
-
-