SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX EXPANDABLE ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C49101320J |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Other, other text: b3: month and year of event have been provided, day is unknown.D4: udi number is unknown; no information has been provided to date.G5: 510k is blank, device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the anesthesia machine did not pass the leak test, and there may be a leak.No adverse patient effects were reported by the customer.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer and h6.Evaluation codes: updated.Email is: (b)(6).One device sample was received without the original package.Per visual inspection, it was not possible to detect any issue which could cause the leaking failure mode.A leak test was performed in which the device passed as no leaks were detected.The complaint was not confirmed.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.No corrective action was taken because the failure mode reported by the customer was not confirmed.
|
|