The most probable root causes associated with this failure mode are software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Per smartphone compatibility information, with use of application, the customer's lg v60 thinq 5g device with android 13 operating system version has not been tested for compatibility at the time of investigation.The compatibility guide states that abbott diabetes care has not assessed compatibility on phone/operating systems other than those listed.This guide is available to the user on the websites where the product is launched.Attempted to activate low glucose alarms using (samsung galaxy s10, android 12, 2.7.2.7077) configuration and successfully received low glucose alarms on the freestyle libre 2 application.Therefore, this issue is not confirmed.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
An alarm issue was reported with the adc device in use with lg v60 thinq 5g phone with os operating system version 13.The low glucose alarm did not sound and customer was not alerted of changes in glucose level.As a result, the customer experienced dizziness, not being able to stay awake, shakiness, confusion, and a loss of consciousness.The customer was provided glucose and juice by a third-party for treatment.There was no report of death or permanent impairment associated with this event.
|