MAUDE Adverse Event Report: SYNTHES GMBH VET K-WIRE Ø1 W/DOUB TIP L150 SST 10U; WIRE, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH VET K-WIRE Ø1 W/DOUB TIP L150 SST 10U; WIRE, SURGICAL Back to Search Results
Catalog Number VW1004.15.10
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative a1, a2, a3, a4, g4 - 510(k): device is a veterinary product.No patient information will be reported.D9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E3: reporter is a j&j employee.H3, h6: the photo investigation revealed that device vw1004.15.10, vet k-wire ø1 w/doub tip l150 sst 10u the pins appear to be slightly bent and also the plastic packaging can be seen in compressed condition.Photographic evidence of the plastic packaging was attached in order to observe in what condition it was received.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the device vw1004.15.10, vet k-wire ø1 w/doub tip l150 sst 10u would contribute to the complained device issue.Based on the investigation findings, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in sweden as follows: it was reported that on an unknown date, at the warehouse it was noticed that the packaging was fully compressed and the pins have bent.This report involves one (1) vet k-wire ø1 w/doub tip l150 sst 10u.This is report 1 of 2 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H6: photo investigation: the product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that device vw1004.15.10, vet k-wire ø1 w/doub tip l150 sst 10u the pins appear to be slightly bent and also the plastic packaging can be seen in compressed condition.Photographic evidence of the plastic packaging was attached in order to observe in what condition it was received.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the device vw1004.15.10, vet k-wire ø1 w/doub tip l150 sst 10u would contribute to the complained device issue.Based on the investigation findings, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H6: device history record (dhr) review conducted: part:vw1004.15.10, lot:7520p74, manufacturing site: werk selzach, supplier:(b)(4), release to warehouse date:22 aug 2023, expiration date: na.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VET K-WIRE Ø1 W/DOUB TIP L150 SST 10U
Type of Device
WIRE, SURGICAL
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach CO 2545
SZ   2545
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key17860665
MDR Text Key324796980
Report Number8030965-2023-12465
Device Sequence Number1
Product Code LRN
Combination Product (y/n)N
Reporter Country CodeSW
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberVW1004.15.10
Device Lot Number7520P74
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received04/08/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/22/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-