MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY RF DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY RF DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number PM2240
Device Problems Premature Discharge of Battery (1057); Failure to Interrogate (1332); No Pacing (3268)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/12/2023
Event Type  Injury  
Event Description
It was reported that patient presented in clinic for routine follow-up.Upon evaluation, it was found that patient's pacemaker was not able to interrogate and there was no magnet response from the pacemaker.No low voltage output and premature battery depletion was also noted.The pacemaker was explanted and replaced.The patient was stable.
 
Manufacturer Narrative
The reported events of no output and no magnet response were confirmed.Premature battery depletion was not confirmed.The device was received with no telemetry communication.Visual inspection of the header attachment area detected an anomaly.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.The device was cut open to enable further testing and the battery was found in normal range.Feedthrough leak test was performed, indicating a device hermeticity breach.This is consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to internal electronics.Hybrid circuitry was tested, and the results indicated normal current drain.A manufacturing process anomaly may have occurred, which resulted in the header bonding anomaly.The device is included in the assurity and endurity pacemakers header anomaly advisory issued by abbott on 15 march 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY RF DR
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17862616
MDR Text Key324822592
Report Number2017865-2023-47620
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507073
UDI-Public05414734507073
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 11/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2017
Device Model NumberPM2240
Device Lot NumberA000024438
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/12/2023
Initial Date FDA Received10/03/2023
Supplement Dates Manufacturer Received10/30/2023
Supplement Dates FDA Received11/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/20/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1530-2021
Patient Sequence Number1
Treatment
ATRIAL LEAD.; RIGHT VENTRICULAR LEAD.
Patient Outcome(s) Required Intervention;
Patient Age90 YR
Patient SexMale
-
-