The logbook provided was analyzed for the investigation.According to this, ventilation stopped on the day of the reported event.An internal leak in the breathing system was identified as the root cause and the vacuum pump was therefore unable to generate the required negative pressure.In accordance with the specification, the device alarmed with "error in the breathing system".In case the root cause cannot be eliminated by the user in this case, an emergency manual resuscitation bag must be used in accordance with the atlan instructions for use, as also described by the user.Technical problems with the vacuum pump, the pump control or leaks in the vacuum branch may be the cause of the fault.The vacuum pump was replaced and examined, but the pressure drop could not be reproduced and no technical fault could be detected.However, the device has been back in operation since november 21 without any further events as described.We therefore assume that the leakage was solved during the replacement of the vacuum pump.Furthermore, therapy mode changes were logged during the corresponding event.From this we can conclude that other ventilation modes were possible, contrary to the user's description.The exact root cause could not be identified.The device was successfully tested on site in accordance with the manufacturer's specifications and put back into operation.
|