DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem No Device Output (1435)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/26/2023
Event Type  malfunction  
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) printer not working.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Manufacturer Narrative
Corrected data: b5, b6, b7, d5, d10, e4, g2, h6 (clinical and impact code).Updated data: b3, b4, e1 (initial reporter and email), e2, e3, g3, g6, h2, h10, h11.
 
Event Description
It was reported that during preventative maintenance by getinge field service engineer, the cs300 intra-aortic balloon pump (iabp) printer not working.There was no patient involvement.
 
Event Description
N/a.
 
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate the unit.The (fse) reported that after replacement of recorder and pcb recorder interface the issue was resolved.The unit was given to customer having passed all functional and safety tests.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key17867882
MDR Text Key324878521
Report Number2249723-2023-04323
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 09/26/2023
Initial Date FDA Received10/04/2023
Supplement Dates Manufacturer Received10/05/2023
02/05/2024
Supplement Dates FDA Received10/09/2023
02/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/18/2017
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.; UNKNOWN.
-
-