• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number R5C4479C
Device Problem Air/Gas in Device (4062)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2023
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a patient experienced air in the line of two (2) homechoice cassettes without an alarm.This occurred during an unspecified process step of peritoneal dialysis therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The devices were not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17868235
MDR Text Key324886124
Report Number1416980-2023-05007
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00085412088433
UDI-Public(01)00085412088433
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberR5C4479C
Device Lot NumberH22H28016
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/07/2023
Initial Date FDA Received10/04/2023
Supplement Dates Manufacturer Received10/20/2023
Supplement Dates FDA Received11/01/2023
Date Device Manufactured08/28/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-