• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH S5 ROLLER PUMP 150 (S5 RP150); CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH S5 ROLLER PUMP 150 (S5 RP150); CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/05/2023
Event Type  malfunction  
Event Description
Livanova deutschland received a report that, during a procedure, the s5 roller pump 150 was not performing its function as a sucker pump.There is no report of any patient injury.
 
Manufacturer Narrative
A.1.-a.5.Patient information were not provided.H10: livanova deutschland manufactures the s5 roller pump.The incident occurred in united states.According to follow up with the customer, the occlusion was not an issue.The perfusionist replace the roller pump with a back-up pump and request a functional inspection be done on the roller pump.A livanova field service engeneer was dispatched to the customer site and connected the complained roller pump to s5 unit to perform functional verification checks per guidelines.However, no error or malfunction of roller pump could be reproduced.Pump was returned to customer and back into service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP 150 (S5 RP150)
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key17868485
MDR Text Key325365725
Report Number9611109-2023-00491
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier04033817900382
UDI-Public(01)04033817900382(11)170725
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K210130
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date 09/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/25/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-