LIVANOVA DEUTSCHLAND GMBH S5 ROLLER PUMP 150 (S5 RP150); CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 10-80-00 |
Device Problem
Pumping Stopped (1503)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/05/2023 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland received a report that, during a procedure, the s5 roller pump 150 was not performing its function as a sucker pump.There is no report of any patient injury.
|
|
Manufacturer Narrative
|
A.1.-a.5.Patient information were not provided.H10: livanova deutschland manufactures the s5 roller pump.The incident occurred in united states.According to follow up with the customer, the occlusion was not an issue.The perfusionist replace the roller pump with a back-up pump and request a functional inspection be done on the roller pump.A livanova field service engeneer was dispatched to the customer site and connected the complained roller pump to s5 unit to perform functional verification checks per guidelines.However, no error or malfunction of roller pump could be reproduced.Pump was returned to customer and back into service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|