MEDTRONIC HEART VALVES DIVISION EVOLUT PRO PLUS DCS; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
|
Back to Search Results |
|
Model Number D-EVPROP23-29 |
Device Problems
Difficult to Remove (1528); Failure to Advance (2524); Device Damaged by Another Device (2915)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/20/2023 |
Event Type
malfunction
|
Event Description
|
Medtronic received information that during the implant of this transcatheter bioprosthetic valve, a non-medtronic guidewire (safari) was set in the left ventricle (lv).A sentrant 14 fr sheath was used to prepare the femoral artery, along with various delivery catheter system (dcs) guides without any problems.The dcs was then used with the non-medtronic guidewire, and felt strange to the physician, as the wire was exiting out the rear of the catheter.According to the physician, the guidewire was not kinked.The physician tried to forcefully advance the dcs in the femoral artery, but it was impossible to move forward or backward.The physician made the decision to cut the wire in order to remove the dcs from the patient.In order to retrieve the dcs out of the femoral artery, the wire was cut just ahead of the nosecone of the dcs.The dcs and wire were then retrieved out of the patient.The other portion of the wire remained in the body.The remaining wire was held with surgical "pince".A procedural delay due to the wire and dcs issue was reported, however the duration of the delay was not reported.The physician once again inserted the external sheath (sentrant), and replaced the wire, dcs and valve to complete the procedure.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Continuation of d10: product id evproplus-23; product lot/serial number (b)(6) ; product type: heart valve; implant date never implanted; product id safari guidewire; product lot/serial number unknown; product type: guidewire product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Updated data: b5 - event description medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received that during the implant of the valve, the non-medtronic guidewire (safari) was cut between the skin and the delivery catheter system (dcs) as they remained out of the body together.The introducer sheath was reinserted on the non-medtronic guidewire that stayed in the patient.The non-medtronic guidewire was then exchanged for a new guidewire.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received that the surgical "pince" referred to was a surgical power scissors.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|