Catalog Number C45101340-NLJ
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/12/2023
Event Type  malfunction  
Event Description
It was reported that during pre-use testing an air leak was found.No harm or adverse effects reported.
 
Manufacturer Narrative
Other, other text: d4: udi section is blank, no information has been provided to date.G5: 510k is blank, device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
G1,2 email is: regulatory.Responses@icumed.Com.Two photos were included for evaluation.One device was received without the original package.Per visual inspection, it was not possible to detect any issue which could cause the leaking failure mode.Per functional testing, the leaking issue was not detected in the leak test; complaint was not confirmed.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.No action was taken since the customer was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ANESTHESIA BREATHING CIRCUIT
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17871222
MDR Text Key324937944
Report Number3012307300-2023-09537
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC45101340-NLJ
Device Lot Number4370866
Was Device Available for Evaluation? No
Date Returned to Manufacturer09/20/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/14/2023
Initial Date FDA Received10/04/2023
Supplement Dates Manufacturer Received11/27/2023
Supplement Dates FDA Received12/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-