• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOQUIP, LLC CARDIOQUIP MODULAR COOLER HEATER; CARDIOPULMONARY BYPASS DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOQUIP, LLC CARDIOQUIP MODULAR COOLER HEATER; CARDIOPULMONARY BYPASS DEVICE Back to Search Results
Model Number MCH-1000(I)
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2023
Event Type  malfunction  
Event Description
Upon inspection of the internal components/tank, our technician identified potential contamination with the unit.The technician took pictures of the internal tubing of the device and forwarded them to cq for review.Based on our findings, epidemiology recommends performing heterotrophic plate count (hpc) testing on the impacted unit to provide a quantitative assessment of water quality.
 
Manufacturer Narrative
Photos of tubing were submitted by a cardioquip technician and sent to cardioquip epidemiology for investigation during an onsite service visit.Due to the discoloration of the tubing, epidemiology recommended that the device receive hpc testing to provide a quantitative assessment of the water quality.The customer was notified of the potential contamination and recommendation via email on 09/12/23.As of the date of this report, there has been no response to the recommendation.A follow-up will be filed if any additional information or information that corrects this report is obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOQUIP MODULAR COOLER HEATER
Type of Device
CARDIOPULMONARY BYPASS DEVICE
Manufacturer (Section D)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer (Section G)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer Contact
charley ford
8422 calibration ct.
college station, TX 77845
MDR Report Key17872071
MDR Text Key324935828
Report Number3007899424-2023-00387
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102147
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberMCH-1000(I)
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/04/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/14/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-