• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 NICORE NITINOL GUIDEWIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 NICORE NITINOL GUIDEWIRE Back to Search Results
Catalog Number 150NSS35
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2023
Event Type  malfunction  
Event Description
It was reported that the tip of the guide wire was broken.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Manufacturer Narrative
The reported event is confirmed cause unknown.Visual evaluation noted the returned sample was found to have the tip of the guide wire bent/broken.This observation of the bent/broken tip does not signify a failure mode of the manufacturing processes.Although an exact root cause could not be determined a potential root cause could be material selection.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "warning: inspect all guidewires for damage prior to use.Bending or kinking during or prior to placement could damage the guidewire.Do not attempt to use the guidewire if it has been damaged.Use of a damaged wire may result in damage to the urinary tract.Do not reshape the guidewire by any means.Attempting to reshape the guidewire may cause damage resulting in release of fragments into the urinary tract.Failure to exercise proper caution may result in damage to the urinary tract.Do not manipulate or remove the guidewire through a metal cannula or needle.This may result in destruction/separation of the outer jacket of the wire requiring retrieval.Failure to comply with the warnings could result in damage to the guidewire to include, but not limited to: wire breakage, abrasion of the coating, release of guidewire fragments into the urinary system, all of which might require intervention." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event Description
It was reported that the tip of the guide wire was broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NICORE NITINOL GUIDEWIRE
Type of Device
NICORE NITINOL GUIDEWIRE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key17877159
MDR Text Key324998683
Report Number1018233-2023-07201
Device Sequence Number1
Product Code EZB
UDI-Device Identifier00801741051524
UDI-Public(01)00801741051524
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number150NSS35
Device Lot NumberNGGP2492
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/11/2023
Initial Date FDA Received10/05/2023
Supplement Dates Manufacturer Received12/05/2023
Supplement Dates FDA Received12/11/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/23/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-