• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG COBAS® SARS-COV-2 & INFLUENZA A/B; COVID-19 MULTI-ANALYTE RESPIRATORY PANEL NUCLEIC ACID DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG COBAS® SARS-COV-2 & INFLUENZA A/B; COVID-19 MULTI-ANALYTE RESPIRATORY PANEL NUCLEIC ACID DEVICES Back to Search Results
Catalog Number 09211101190
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/20/2023
Event Type  malfunction  
Event Description
In light of the covid-19 pandemic and the subsequent emergency use authorizations (euas) for sars-cov-2 diagnostic tests, the agency has requested heightened reporting beyond the reasonably suggests requirements of 803 to include allegations of false positive or false negative results independent of harm or malfunction or off-label use.Pursuant to the agency¿s instruction, we hereby submit this mdr.A customer from australia alleged discrepant results for one patient while using the cobas® sars-cov-2 & influenza a/b nucleic acid test on the cobas® liat® system.The alleged sample initially generated a negative result for all targets (sars-cov-2 and influenza a&b).The same sample was retested on another assay (genexpert sars-cov-2/flu/rsv plus) and generated a positive result for sars-cov-2.The same sample was stored overnight and then retested on a different liat analyzer which was negative for all targets, and the genexpert sars-cov-2/flu/rsv plus assay which was positive for sars-cov-2 the initial negative result was reported and then amended.No harm was alleged.An investigation was conducted to evaluate the customer issue.Per fda¿s eua guidance, 1 mdr will be filed.
 
Manufacturer Narrative
For the retest on 21-sep-2023, it was observed that the sample was stored improperly overnight which may have led to some degree of sample degradation.Based on the data analysis review, the product is performing as intended.A result discrepancy may be expected between assays due to potential differences in the tests¿ limits of detection (lod) and intended use.Note that not following the method sheet for sample storage likely caused sample degradation.---- lot 30103g expiration date: 6/30/2024.Lot 21205f expiration date: 5/31/2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS® SARS-COV-2 & INFLUENZA A/B
Type of Device
COVID-19 MULTI-ANALYTE RESPIRATORY PANEL NUCLEIC ACID DEVICES
Manufacturer (Section D)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
branchburg NJ 08876
Manufacturer (Section G)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
branchburg NJ 08876
Manufacturer Contact
timothy blair
1080 us hwy 202 s
branchburg, NJ 08876
MDR Report Key17883871
MDR Text Key325077526
Report Number2243471-2023-03208
Device Sequence Number1
Product Code QLT
Combination Product (y/n)Y
Reporter Country CodeAS
PMA/PMN Number
EUA201779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number09211101190
Device Lot Number30103G,21205F
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/29/2023
Initial Date FDA Received10/06/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-