• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK DUO-VENT CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK DUO-VENT CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C8541
Device Problems Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/10/2023
Event Type  malfunction  
Manufacturer Narrative
G1: device manufacturer address 1:(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a clearlink system continu-flo solution set had a hole in the line under the drip chamber resulting in a leak; further described as normal saline "began to flow and spray out the very top of the line".This was identified during priming.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: the device was received for evaluation.A visual inspection was performed which observed a twist at the joint of the tubing and drip chamber.Functional tests including pressure test and clear passage tests were performed and the results were not satisfactory; there was a leak at the joint of the tubing and the drip chamber.The reported condition was verified.The cause of the condition was a manufacturing related issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK DUO-VENT CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17886726
MDR Text Key325116162
Report Number1416980-2023-05070
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412071107
UDI-Public(01)00085412071107
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K203609
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2C8541
Device Lot NumberR23E27109
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received11/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-