Model Number CDHFA500Q
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2023
Event Type  malfunction  
Event Description
It was reported during an unrelated procedure on (b)(6) 2023 the set screw for the atrial port was not working properly.The implantable cardioverter defibrillator (icd) was removed and replaced within the same operation.Post-procedure the patient was stable.
 
Manufacturer Narrative
The reported event of set screw anomaly was confirmed.The device was above the elective replacement indicator (eri) upon receipt.Analysis revealed the atrial set screw was unseated.This prevented full insertion of the torque driver and was the cause of the reported event.The set screw anomaly was consistent with having occurred during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17887472
MDR Text Key325124649
Report Number2017865-2023-47958
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Model NumberCDHFA500Q
Device Lot NumberS000083954
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/22/2023
Initial Date FDA Received10/06/2023
Supplement Dates Manufacturer Received12/21/2023
Supplement Dates FDA Received01/10/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/14/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ATRIAL LEAD; LEFT VENTRICULAR LEAD; RIGHT VENTRICULAR (RV) LEAD
Patient Age80 YR
Patient SexMale
-
-