Model Number M00568211
Device Problem Entrapment of Device (1212)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/13/2023
Event Type  malfunction  
Event Description
Note: this report pertains to the second of three endovive standard peg kit push used during the same procedure.It was reported to boston scientific corporation that an endovive standard peg kit push method was used during an esophagogastroduodenoscopy procedure performed on (b)(6) 2023.It was reported that when attempting to push the peg tube over the wire, it kept getting stuck where the introducer meets the peg tube.After multiple attempts to remove the peg tube which was stuck on the wire.The wire became unraveled and stretched.The device was removed from the patient.A second peg kit push method was opened however the same thing happened.The physician then used an endovive standard pull kit to complete the procedure.After the procedure, a third endovive standard peg push tube with the same lot was opened and the same thing occurred with the ones used previously.There were no patient complications reported as a result of this event.
 
Manufacturer Narrative
Block h6 (device codes): imdrf device code a1409 captures the reportable event of peg tube obstructed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOVIVE STANDARD PEG KIT
Type of Device
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17890227
MDR Text Key325143156
Report Number3005099803-2023-05364
Device Sequence Number1
Product Code KNT
UDI-Device Identifier08714729285205
UDI-Public08714729285205
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K031538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00568211
Device Catalogue Number6821
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-