• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS, 2D; FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS, 2D; FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM Back to Search Results
Model Number SDM-00001-2D
Device Problem Unintended System Motion (1430)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/19/2023
Event Type  malfunction  
Event Description
It was reported that on (b)(6) 2023, during a selenia dimensions procedure the c-arm kept rotating past 45 degree angle.A field engineer examined the equipment and replaced both switch decks and paddle switch assembly.Upon installation the right switch deck did not performed as expected.A faulty connection was observed and reassembled.The system met the manufacturer´s specifications.No patient injury reported.No other information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS, 2D
Type of Device
FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
36 & 37 apple ridge road
danbury CT 06810
Manufacturer (Section G)
HOLOGIC, INC.
36 & 37 apple ridge road
danbury CT 06810
Manufacturer Contact
daniel guevara
562 parkway
coyol free zone building b24
san jose 20102
CR   20102
MDR Report Key17890476
MDR Text Key325380582
Report Number1220984-2023-00070
Device Sequence Number1
Product Code MUE
UDI-Device Identifier15420045505582
UDI-Public(01)15420045505582(11)160824
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSDM-00001-2D
Device Catalogue NumberSDM-00001-2D
Was Device Available for Evaluation? No
Date Manufacturer Received09/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/24/2016
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-