Model Number CI512
Device Problems Failure to Deliver Shock/Stimulation (1133); Failure to Read Input Signal (1581); Output Problem (3005); Therapeutic or Diagnostic Output Failure (3023)
Patient Problems Failure of Implant (1924); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Type  Injury  
Manufacturer Narrative
This report is submitted on october 9, 2023.
 
Event Description
Per the clinic, the patient experienced a loss of connection to the internal device subsequent to sustaining a head trauma.Troubleshooting at the clinic did not resolve the issue and the clinic requested a cochlear specialist to verify the device function.There are plans to explant the device and to reimplant the patient with a new device; however, this has not occurred as of the date of this report.
 
Manufacturer Narrative
Per the clinic, the device was explanted on (b)(6) 2023 and the patient was reimplanted with another cochlear device during the same surgery.This report is submitted on november 20, 2023.
 
Manufacturer Narrative
Device analysis report attached.This report is submitted on january 09, 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS CI512 COCHLEAR IMPLANT WITH CONTOUR ADVANCE ELECTRODE
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LIMITED
1 university avenue
macquarie university
nsw 2109
AS  2109
Manufacturer (Section G)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS   2109
Manufacturer Contact
thevamalar jayamagentharan
unit ug-1, vertical podium
no. 8 jalan kerinchi,
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key17892749
MDR Text Key325232126
Report Number6000034-2023-03271
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09321502014564
UDI-Public(01)09321502014564(11)110318(17)130317
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/08/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date03/17/2013
Device Model NumberCI512
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-