Related manufacturer reference number: 2017865-2023-48058.It was reported that the patient presented for implantable cardioverter defibrillator (icd) implant.During the procedure, it was noted that ventricular pacing inhibition occurred upon connection of the leads to the header, resulting in patient asystole.It was further noted that the left ventricular (lv) lead was suspected to have micro-dislodged.Upon reconnection of the leads, it was still noted that there was no low voltage output from the icd.The procedure was completed using an alternate lv lead and icd on (b)(6) 2023.The patient was stable.
|
The reported event of an output, pacing, and header anomaly was not confirmed.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.No anomalies were found.
|