BOSTON SCIENTIFIC CORPORATION EASYTRAK 2; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number 4518 |
Device Problems
Failure to Capture (1081); Low impedance (2285); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/26/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that electrogram (egm) review of this cardiac resynchronization therapy defibrillator (crt-d) system revealed atrial undersensing and intermittent left ventricular (lv) lead loss of capture (loc).A device check from march 2023 showed a lv threshold measurement of 2.8v @1ms and lv output set to 3.5v@1ms.Additionally, all lead impedance measurements started trending downward in early (b)(6) 2023 that correlated with an upward trend in the patient's heartlogic index, which suggests that the decrease in all impedance measurements was likely attributed a possible heart failure and/or fluid retention.Technical services (ts) reviewed troubleshooting options and recommended further evaluation in the clinic.This crt-d system remains in service.No adverse patient effects were reported.
|
|
Search Alerts/Recalls
|
|
|