• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE BD INSULIN SYRINGES WITH BD ULTRA-FINE¿ NEEDLE; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE BD INSULIN SYRINGES WITH BD ULTRA-FINE¿ NEEDLE; PISTON SYRINGE Back to Search Results
Catalog Number 328418
Device Problem Break (1069)
Patient Problem Foreign Body In Patient (2687)
Event Date 09/25/2023
Event Type  malfunction  
Event Description
It was reported that the bd insulin syringes with bd ultra-fine¿ needle broke off.The following information was provided by the initial reporter: consumer reported found 1 syringe needle broke off in her arm.She was able to remove needle from arm herself.No medical attention needed.Caller denied reuse of needles.Long time diabetic.Item in her sharp's container.
 
Manufacturer Narrative
B.3.Date of event: unknown.The date received by manufacturer has been used for this field.H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Manufacturer Narrative
H.6.Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
 
Event Description
It was reported that the bd insulin syringes with bd ultra-fine¿ needle broke off.The following information was provided by the initial reporter: consumer reported found 1 syringe needle broke off in her arm.She was able to remove needle from arm herself.No medical attention needed.Caller denied reuse of needles.Long time diabetic.Item in her sharp's container.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSULIN SYRINGES WITH BD ULTRA-FINE¿ NEEDLE
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17901660
MDR Text Key325481994
Report Number1920898-2023-00687
Device Sequence Number1
Product Code FMF
UDI-Device Identifier00382908418034
UDI-Public(01)00382908418034
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number328418
Device Lot Number2339221
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/25/2023
Initial Date FDA Received10/09/2023
Supplement Dates Manufacturer Received10/20/2023
Supplement Dates FDA Received10/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/05/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-