• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH BATTERY RECIPROCATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH BATTERY RECIPROCATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.715
Device Problems Device-Device Incompatibility (2919); Mechanical Jam (2983); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that during service and evaluation, it was determined that the battery reciprocator device would not run - trigger defective and sticky trigger.It was further determined that the device failed pretest for general condition, check for sticky, check function of device and check oscillation frequency.It was noted in the service order that the device blade could not be attached and the head would not rotate to allow the blade to attach.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, an additional report will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device was returned for evaluation.During evaluation it was determined that the initial reported condition was not confirmed.Therefore, the assignable root cause was not determined.However, the malfunctions found during service and evaluation have been confirmed.The assignable root cause was determined to be due to component wear.Ud: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY RECIPROCATOR II FOR BPL II
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key17905131
MDR Text Key325292156
Report Number8030965-2023-12792
Device Sequence Number1
Product Code GEY
UDI-Device Identifier07611819491854
UDI-Public07611819491854
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.715
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2023
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/09/2023
Initial Date FDA Received10/10/2023
Supplement Dates Manufacturer Received10/09/2023
Supplement Dates FDA Received10/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-