DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 998-00-3023-55 |
Device Problem
Failure to Run on Battery (1466)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/03/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that, the cs300 intra-aortic balloon pump (iabp) unit not working on battery backup.
|
|
Event Description
|
It was reported that before patient use during checks by customer, the cs300 intra-aortic balloon pump (iabp) unit not working on battery backup.There was no patient involvement.
|
|
Manufacturer Narrative
|
Updated fields: (type of investigation, investigation findings, component, investigation conclusions), h10.A getinge field service engineer (fse) was dispatched to evaluate the iabp unit and the fse found that the machine is showing alarm low battery in spite of the batteries fully charged, cross checked with other working battery pack also.The charging voltages from the power supply is ok and the fuses are also ok, but still the machine is showing false alarm.Also the charging indication is not showing.Needed power supply for testing purpose.The power supply was then replaced, now the charging and battery indication both are working.All functional tests are passed.Machine is working ok.The fse handed over equipment to customer in good working condition.
|
|