DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 998-00-3023-55
Device Problem Failure to Run on Battery (1466)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/03/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that, the cs300 intra-aortic balloon pump (iabp) unit not working on battery backup.
 
Event Description
It was reported that before patient use during checks by customer, the cs300 intra-aortic balloon pump (iabp) unit not working on battery backup.There was no patient involvement.
 
Manufacturer Narrative
Updated fields: (type of investigation, investigation findings, component, investigation conclusions), h10.A getinge field service engineer (fse) was dispatched to evaluate the iabp unit and the fse found that the machine is showing alarm low battery in spite of the batteries fully charged, cross checked with other working battery pack also.The charging voltages from the power supply is ok and the fuses are also ok, but still the machine is showing false alarm.Also the charging indication is not showing.Needed power supply for testing purpose.The power supply was then replaced, now the charging and battery indication both are working.All functional tests are passed.Machine is working ok.The fse handed over equipment to customer in good working condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key17905586
MDR Text Key325297199
Report Number2249723-2023-04382
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567112541
UDI-Public10607567112541
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 01/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number998-00-3023-55
Device Catalogue Number998-00-3023-55
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/07/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-