MEDOS INTERNATIONAL SÃ RL CH CONFIDENCE SPINAL CEMENT SYSTEM CONFIDENCE PLUS KIT SPINAL CEMENT SYSTEM 11CC; CEMENT, BONE, VERTEBROPLASTY
|
Back to Search Results |
|
Catalog Number 283910000 |
Device Problems
Chemical Problem (2893); Failure to Eject (4010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2023 |
Event Type
malfunction
|
Event Description
|
Device report from synthes reports an event in portugal as follows: it was reported that on an unknown date, after mixing the confidence cement powder with the solvent, the final mixture ended up too dry and stiff, preventing it from being injected as usual through the cannulas.The procedure was completed successfully with no patient consequence due to the event.This report involves one confidence spinal cement system confidence plus kit spinal cement system 11cc.This is report 2 of 2 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: b3: only the event year is known.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j employee.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The product was not returned to depuy synthes, however photos were provided for review.The photo investigation found nothing that could be related to a product malfunction.Retained sample test was performed by the manufacturer and the product behaved as intended, for full report please refer to attachment "(b)(4) report".As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was unconfirmed as the observed evidence doe snot represent the reported allegation with the [confidence spinal cmt sys, 11c].Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.A manufacturing record evaluation was performed for the finished device product code#:283910000 lot #:371514 it was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on:15/05/2023.Manufacturing site: jabil le locle.Expiry date:30/04/2025.Cement number: product code :183901001 / batch number : 4071985.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|