MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ATTUNE RP PS ARTIC SURF SZ8; ATTUNE INSTRUMENTS : TIBIAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ATTUNE RP PS ARTIC SURF SZ8; ATTUNE INSTRUMENTS : TIBIAL TRIALS Back to Search Results
Catalog Number 254500548
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/30/2023
Event Type  malfunction  
Manufacturer Narrative
Product complaint # :(b)(4).Investigation summary: black spots all over.The device associated with this report was returned to depuy synthes for evaluation.Cleaning/sterilization protocols were asked, however they were not provided.Visual analysis of the returned sample revealed that attune rp ps artic surf sz8 confirmed the presence of black particles firmly adhered to the surface of the device.From the observations made of the previous investigation ((b)(4)) as well as from the similar physical observations made of the current complaint sample, the black particle contamination is considered not related to material or manufacturing defects.Additionally, the device is chipped at the anterior aspect.The observed condition of the device was consistent with a component failure that may have been caused by exposure to unintended forces with another tool.The broken fragment was not returned.The overall complaint was confirmed as the observed condition of the attune rp ps artic surf sz8 would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to maintenance and unintended use error.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H6: component code: appropriate term/code not available (g07002) used to capture no findings available.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that there were black spots all over the instrument.The product damage documented in this pc has been identified during loan kit inspection, by the loan kit technician.The event date and alert date are the date that the inspection took place.There is no surgeon, procedure, or patient details available.No further information can be obtained as the case was not reported by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP PS ARTIC SURF SZ8
Type of Device
ATTUNE INSTRUMENTS : TIBIAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key17905987
MDR Text Key325300551
Report Number1818910-2023-20579
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295132882
UDI-Public10603295132882
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 10/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500548
Device Lot NumberMVMCJT990
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/08/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberZ-2139-2015
Patient Sequence Number1
-
-