Model Number CD3371-40QC
Device Problems Over-Sensing (1438); Pacing Asynchronously (1441)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/26/2023
Event Type  malfunction  
Manufacturer Narrative
Session records were provided for review by technical services.Analysis of the session records indicated that the post-paced t-wave oversensing (pptwos) associated with fusion event was sensed by the device.
 
Event Description
It was reported that during a follow up in clinic, post-paced t-wave oversensing (pptwos) was noted on the device.Abbott technical support was contacted and the pptwos was suspected to be due to fusion at the maximum tracking rate (mtr).The device was reprogrammed to resolve the event.The patient was in stable condition.Ble.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
No Match
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17908539
MDR Text Key325340479
Report Number2017865-2023-47439
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2021
Device Model NumberCD3371-40QC
Device Lot NumberP000077080
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/26/2023
Initial Date FDA Received10/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/08/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-