Calibration data was acceptable.Liquid flow cleaning was last performed on (b)(6) 2023.A serum and plasma sample (same draw) from the patient was submitted for investigation.Elecsys hiv combi results: serum sample: 0.694 coi (negative).Plasma sample: 0.783 coi (negative).These results are considered to be elevated negative results.Further investigation showed both samples were positive with one hiv-1 specific antigen of the elecsys hiv combi assay.The serum sample was tested with biorad new lav blot i, the biorad new lav blot ii, and the fujirebio inno-lia hiv i/ii score.The sample showed one faint band (p24) in the biorad hiv-1 blot (indeterminate) and showed no band in the biorad hiv-2 blot (negative).These results lead to an indeterminate hiv-1 blot result.In the inno-lia hiv i/ii score the sample showed no band which leads to a negative result interpretation.Based on the investigation results, the patient sample was likely drawn during the very early seroconversion phase.In this early phase of infection, the hiv analyte concentration could be very low which may lead to negative results.A negative test result does not completely rule out the possibility of an infection with hiv.Serum or plasma from the very early phase of hiv infections can occasionally yield negative findings.For diagnostic purposes, the results should always be assessed in conjunction with the patient's medical history, clinical examination, and other findings.The elecsys hiv combi pt performs within specification.The investigation did not identify a product problem.
|