Catalog Number SGC0702 |
Device Problems
Break (1069); Positioning Failure (1158)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/18/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device will not be returned for evaluation as the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
This is filed to report a cable break it was reported that during preparation of the steerable guide catheter (sgc), while applying minus knob, the physician felt a snap in the guide.The sgc was not longer able to straighter or curve; therefore, it was not used and was replaced.There was no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
The device was not returned.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no similar complaints from the lot.All information was investigated and without the device to analyze, a cause for the reported cable break could not be determined.The reported positioning failure of unable to curve and straighten the tip and noise (audible) however, appear to be cascading effects of the cable break.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
|