• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2023
Event Type  malfunction  
Event Description
Philips received a complaint from a philips medical device maintenance specialist (mdms) reporting that the battery on the v60 ventilator could not charge.The device was not in clinical use.There was no report of harm.There was no battery failed alarm and the battery was older than 5 years of age.Per the philips v60 service manual, recommended timetable for battery replacement is every 5 years.The battery replacement is based on the date of manufacture recorded on the battery label.Also, it was reported that the customer stored the unit without being connected to an alternating current (ac) power source.The v60 service manual advises the user the battery may fail to charge if stored for an extended period of time without the ventilator plugged into ac power.This issue was caused by use error.Per gfe response, it was confirmed that the battery will be replaced to resolve the reported issue.Once replaced, the required performance verification tests per philips standards will be performed and the device will be returned to service.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Manufacturer Narrative
E1 reporter phone number - (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key17914741
MDR Text Key325417434
Report Number2518422-2023-26331
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838009844
UDI-Public00884838009844
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60
Device Catalogue Number1053613
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/05/2023
Initial Date FDA Received10/11/2023
Date Device Manufactured08/19/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-