Model Number G247
Device Problems Inappropriate/Inadequate Shock/Stimulation (1574); Inaccurate Synchronization (1609)
Patient Problem Electric Shock (2554)
Event Date 10/04/2023
Event Type  Injury  
Event Description
It was reported that, this cardiac resynchronization therapy defibrillator (crt-d) delivered six inappropriate shocks due to atrial arrythmia.The episode began with an atrial tachycardia.Then, the device charged but the rhythm broke, and the shock was diverted.The rhythm re initiated five seconds later and met redetection, the device charged and committed the shock due to prior divert, despite rhythm slowing.This pattern of short bursts with charge diverted followed by committed shock continues for a total of six shocks delivered and there were parts of acceleration.The technical services (ts) discussed programming options.This device remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIGILANT X4 CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17916942
MDR Text Key325444375
Report Number2124215-2023-56417
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589287
UDI-Public00802526589287
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/29/2021
Device Model NumberG247
Device Catalogue NumberG247
Device Lot Number223973
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/10/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexMale
-
-