COLOPLAST A/S X-FLOW PROSTATECTOMY CATHETER; INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL Back to Search Results
Catalog Number AB6318
Device Problems Burst Container or Vessel (1074); Material Rupture (1546)
Patient Problem Insufficient Information (4580)
Event Date 08/23/2023
Event Type  malfunction  
Manufacturer Narrative
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Event Description
According to available information, this device had a balloon crack.The patient needed to be catheterized again.No other adverse patient effects were reported.
 
Manufacturer Narrative
After receiving this complaint, we searched for other complaint and we didn¿t find other complaint on the lot number 9282506.Checking the quality databases revealed one non-conformity in relation to the described defect and a corrective and preventive action: - nc (b)(4): "pb ballons (bulles + agglutinés)" opened in 16 january 2023.For this nc, the used of clumped balloons should be responsible of some weakness on the balloon and being responsible of bursting.- monitoring capa-000152: "balloon issues on folysil and silicone prostatic catheters".The trending for balloon bursting is specifically monitored.On 14th december, we received documentary investigation by our subcontractor which mentionned known issue with raw material balloon, a similar case study based on item number ab63xx, defect balloon burst, over last four year: 55 similar case were found.
 
Event Description
According to available information, this device had a balloon crack.The patient needed to be catheterized again.No other adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-FLOW PROSTATECTOMY CATHETER
Type of Device
INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer (Section G)
CMF-SARLAT
9 avenue edmond rostand
sarlat-la-caneda
FR  
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key17917559
MDR Text Key325458290
Report Number9610711-2023-00219
Device Sequence Number1
Product Code EZL
UDI-Device Identifier03600040144183
UDI-Public3600040144183
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberAB6318
Device Lot Number9282506_AB63181002
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-