Model Number WB991046 |
Device Problems
Device Alarm System (1012); Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
The customer reported that when she powers on her olympus electrosurgical generator, an alarm begins to sound, and the unit produces an error code.Additional details relating to the patient and the event have been requested, but no response has been received at this time.There was no patient harm or consequence reported as a result of this event.
|
|
Manufacturer Narrative
|
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user¿s report was confirmed.A defective printed circuit board was discovered and caused the reported alarm and error code (e283).The front mask was cracked with minor scratching on the housing covers.If additional information becomes available following the device evaluation, a supplemental report will be filed.
|
|