LIVANOVA DEUTSCHLAND GMBH S5 DOUBLE ROLLER PUMP 85; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2023
Event Type  malfunction  
Event Description
Livanova deutschland received a report that dual rollerhead pump failure during cardioplegia administration.After cross clamp was placed, cardioplegia pump was turned on to administer arrestingc ardioplegia dose.A clamp was left on cardioplegia line at the field, so high line pressure alarm was triggered on the cardioplegia pump.Clamp at field was opened and high pressure was relieved.When trying to resume arresting cardioplegia dose, the cardioplegia pump failed to move/run.Perfusion cleared the high-pressure alarm and ensured that pump was in the correct delivery mode and even tried to deliver in a recirculation mode.When trying to turn the pump on again, rpm¿s and flows were being shown without the rollerhead moving.To arrest the heart, the perfusionist had to remove the tubing from the pump head and hand squeeze cardioplegia to the field for the surgeon to deliver by syringe.Once heart was arrested, cardioplegia pump was turned off and back on and resumed working.There is no report of any patient injury.
 
Manufacturer Narrative
A.1.-a.5.Patient information were not provided.H10: livanova deutschland manufactures the s5 double roller pump.The incident occurred in united states.A livanova field service engineer was dispatched to the customer site.Evaluation and functional checks were performed and no issue could be reproduced.Unit is functional and ready to use.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Event Description
See initial report.
 
Manufacturer Narrative
H10: the serial read-out of the pump (real time device parameters and setting recording file) was analyzed: no technical error was found on the date of the event excluding a hardware malfunction.Based on technical intervention carried out and considering the serial read-out analysis, it cannot be ruled out that the inability to restart the pump could be caused by any monitoring function activated during the procedure with still alarm condition on (e.G.Pressure alarm).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE ROLLER PUMP 85
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key17921236
MDR Text Key326375924
Report Number9611109-2023-00519
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K210130
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/08/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Date FDA Received10/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/24/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-