ROCHE DIAGNOSTICS ELECSYS FT4 IV; FREE THYROXINE Back to Search Results
Catalog Number 09043276190
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/20/2023
Event Type  malfunction  
Event Description
There was an allegation of questionable thyroid results for one patient from the cobas 6000 e601 module compared to an abbott architect analyzer.Refer to the attachment to the medwatch for all patient data.This medwatch is for the ft4 g4 elecsys assay.Refer to the medwatch with a1 patient identifier (b)(6) for the ft3 g3 elecsys assay.No information was provided to determine if the questionable result was reported outside of the laboratory.
 
Manufacturer Narrative
The cobas 6000 e601 serial number was (b)(6).The investigation is ongoing.
 
Manufacturer Narrative
The anti-tpo result from the cobas 6000 e601 module as part of the investigation was actually 550.4/538.7 iu/ml, not <30 iu/ml.
 
Manufacturer Narrative
Additional data was provided for the patient sample.The sample was sent to an outsourced laboratory for additional testing and was tested for submitted for preliminary investigation by competitor methods.Refer to the attachment to the medwatch for all patient data.Sample from the patient was requested for further investigation.
 
Manufacturer Narrative
Sample from the patient was investigated and an interfering factor to the assay components was detected.Product labeling for the assay states: in rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies, streptavidin or ruthenium can occur.These effects are minimized by suitable test design.For diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination and other findings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 IV
Type of Device
FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17921334
MDR Text Key325506517
Report Number1823260-2023-03277
Device Sequence Number1
Product Code CEC
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
K220456
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number09043276190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/28/2023
Initial Date FDA Received10/12/2023
Supplement Dates Manufacturer Received10/18/2023
10/26/2023
03/15/2024
Supplement Dates FDA Received11/09/2023
11/16/2023
03/15/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age46 YR
Patient SexFemale
-
-