• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES INSPIRIS RESILIA AORTIC VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES INSPIRIS RESILIA AORTIC VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number 11500A
Device Problem Patient-Device Incompatibility (2682)
Patient Problem Dyspnea (1816)
Event Date 09/19/2023
Event Type  Injury  
Event Description
It was reported that a patient with a 25mm 11500a aortic valve has been placed under evaluation for a valve-in-valve after an implant duration of 4 years, 3 months due to prothesis mismatch (ava 0.81 cm2).The patient presented with shortness of breath.
 
Manufacturer Narrative
H10: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Manufacturer Narrative
Patient prosthesis mismatch (ppm) is present when the effective orifice area of the inserted prosthetic valve is too small in relation to body size.Its main hemodynamic consequence is to generate higher than expected gradients through a normally functioning prosthetic valve.The root cause of this event cannot be conclusively determined with the available information; however, patient factors likely caused or contributed to the event.No information was provided to indicate the patient's weight and height prior to implant of the initial surgical valve.This event does not allege a labeling non-conformance/deficiency, a use related issue with a hazardous situation, or a device related infection and there is no evidence of product failure with regard to design, reliability, or use error.The instructions for use (ifu) have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.The subject device is not available for evaluation as it remains implanted.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRIS RESILIA AORTIC VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2 - office m2013
irvine, CA 92614
9492506615
MDR Report Key17921417
MDR Text Key325507504
Report Number2015691-2023-16737
Device Sequence Number1
Product Code LWR
UDI-Device Identifier00690103194999
UDI-Public(01)00690103194999(17)200214
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/14/2020
Device Model Number11500A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/14/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Life Threatening;
Patient Age89 YR
Patient SexMale
-
-