Catalog Number 07.702.016S |
Device Problems
Break (1069); Improper Chemical Reaction (2952)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/25/2023 |
Event Type
malfunction
|
Event Description
|
Device report from synthes reports an event in malaysia as follows: it was reported that the blue handle which was supposed to be pushed and turned to mix the cement inside the tube, was not moving.When we try to twist and push pull the blue handle to mix the cement , it got broken off.The handle got broken and separated from the tube causing the cement not mixed properly and hardens inside the tube.Therefore we immediately open a new set so as not to delay the operation.The surgery was delay for less than 5 minutes.The procedure was successfully completed.No fragments were generated.There were no patient outcome or consequences.This report is for one (1) vertecem v+ cement kit.This is report 1 of 1 for complaint (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D9: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
After mixing saline into powder, use the blue handle and push up and down to mix, but it does not move.When the surgeon tried to push the blue handle, the blue handle broke.The cement was not mixed yet.The blue handle used for mixing broke when trying to mix the saline with powder.The cement was not mixed yet.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.G4: device is not distributed in the united states, but is similar to device marketed in the usa.The product was not returned to depuy synthes, however photo was provided for review.See attachment (product complain-vertecem cement).The photo investigation revealed that device 07.702.016s, vertecem v+ cement kit had broken blue handle which is separated from the tube and based on the available evidence it is not possible to confirm the allegation cement will not mix properly and cement setting time.The cement retain sample test was requested for the finished device product code - 07.702.016s, lot code - 2l53560 and it shows no deviations.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the device 07.702.016s, vertecem v+ cement kit would/would not contribute to the complained device issue.Based on the investigation findings, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Part: 07.702.016s.Lot: 2l53560.Release to warehouse date: 03 nov 2022.Manufacturing site: werk selzach.Expiration date: 01 nov 2025 supplier: (b)(4).A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|