• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71953-01
Device Problems Premature Discharge of Battery (1057); Unable to Obtain Readings (1516)
Patient Problems Hypoglycemia (1912); Nausea (1970)
Event Date 09/20/2023
Event Type  Injury  
Manufacturer Narrative
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The date of event is unknown.The date entered in section b3 is the date abbott diabetes care became aware of the event.The device mfg date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A battery issue was reported with the adc device.Customer was unable to obtain readings due to a fast draining battery.As a result, customer experienced symptoms of "drowsy and nausea" and had contact with an healthcare professional (hcp) and was treated with fluids for diagnosis of hypoglycemia.Customer stated "they had their sugar, liver, kidney and heart checked" no further treatment was required.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The date of event is unknown.The date entered in section b3 is the date abbott diabetes care became aware of the event.The device mfg date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.This serves as a correction report.
 
Event Description
A battery issue was reported with the adc device.Customer was unable to obtain readings due to a fast draining battery.As a result, customer experienced symptoms of "drowsy and nausea" and had contact with an healthcare professional (hcp) and was treated with fluids for diagnosis of hypoglycemia.Customer stated "they had their sugar, liver, kidney and heart checked" no further treatment was required.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned and a valid serial number has not been provided for reader and cable adapter.Therefore, it is not possible to check dhr for cable and adapter.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.A tripped trend review was conducted for the reported complaint and libre reader, no trends were identified that would indicate any product related issues.If the product is returned, a physical investigation will be performed and a follow up report submitted.All pertinent information available to abbott diabetes care has been submitted.Upon extended investigation, it was determined that the serial number provided by the customer ((b)(6)) and previously reported to the fda was not a valid serial number.Therefore, section d4 was updated to unk.
 
Event Description
A battery issue was reported with the adc device.Customer was unable to obtain readings due to a fast draining battery.As a result, customer experienced symptoms of "drowsy and nausea" and had contact with an healthcare professional (hcp) and was treated with fluids for diagnosis of hypoglycemia.Customer stated "they had their sugar, liver, kidney and heart checked" no further treatment was required.There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key17932029
MDR Text Key325626377
Report Number2954323-2023-45856
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 12/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71953-01
Was Device Available for Evaluation? No
Date Manufacturer Received12/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-