The technical support representative instructed 'carefully' to remove the specimen from the probe.Operator received a small cut to finger through gloved hand during trouble shooting the instrument.The acl top series operator's manual confirmed that there are appropriate instructions and cautions regarding the activities of a stuck or jammed cts piercer probe in the cap of a primary sample tube.Per the acl top family 50 series operator's manual; " warning piercing hazard: do not put hand inside instrument while sample arm is in motion.Caution biohazard: in the unlikely event that the cts piercer probe gets stuck or jammed in the cap of a primary sample tube, do not attempt to remove the cap.Call service for assistance.The cts piercer probe assembly is delicate, very sharp, and biohazardous.The removal of a stuck cap from the piercer probe could cause personal injury or damage to the instrument".No remedial action required.
|