• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT VR; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT VR; NO MATCH Back to Search Results
Model Number CDVRA500Q
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2023
Event Type  malfunction  
Event Description
During an initial implant procedure on (b)(6) 2023, it was reported that the set screw of the implantable cardioverter defibrillator (icd) was too loose and would not accept the lead.The physician tried different wrenches, but it still would not tighten.The icd was not used, and a new icd was successfully implanted.The patient was stable throughout the procedure.
 
Manufacturer Narrative
The reported event of set screw issue was confirmed.The device was brought to the bench and the header was inspected further, and no screw was detected when inserting a test lead into the rv (df4) lead bore.The issue was consistent with having occurred during the procedure and traced to the missing rv (df4) set screw.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT VR
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17935954
MDR Text Key325693745
Report Number2017865-2023-49652
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067031990
UDI-Public05415067031990
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDVRA500Q
Device Lot NumberA000146862
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/13/2023
Supplement Dates Manufacturer Received12/06/2023
Supplement Dates FDA Received12/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA TACHY MRI LEADS
Patient Age67 YR
Patient SexFemale
Patient Weight44 KG
-
-