MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1712K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-1712K
Device Problem Device Alarm System (1012)
Patient Problems Hyperglycemia (1905); Polydipsia (2604)
Event Date 09/20/2023
Event Type  malfunction  
Manufacturer Narrative
This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Select patient information cannot be provided due to regional privacy regulations.The reported device is not marketed in the united states, but it is a same/similar device to one that is marketed outside the united states.Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Information received by medtronic indicated that the customer experienced hyperglycemia and the pump gives an alarm and does not allow it to be filled any further.Troubleshooting was performed but the issue was not resolved.No harm requiring medical intervention was reported.The customer will discontinue to use the insulin pump and the insulin pump will be returned for analysis.
 
Manufacturer Narrative
The pump passed the displacement test, self-test, and p-cap locks properly into the reservoir compartment.Successfully downloaded the trace and history files using thus and carelink.Pump was cut open to perform visual inspection and found no evidence of physical or moisture damage on the electronic assembly, motor, or force sensor.No absence of alarm anomaly noted during analysis.The following were noted during visual inspection: scratched case, pillowing keypad overlay, corroded battery tube, and stained keypad overlay.No absence of alarm anomaly noted during analysis.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
640G INSULIN PUMP MMT-1712K
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
lusine boyadzhyan
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key17937189
MDR Text Key326388837
Report Number2032227-2023-283514
Device Sequence Number1
Product Code OYC
UDI-Device Identifier000000763000317157
UDI-Public(01)000000763000317157
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1712K
Device Catalogue NumberMMT-1712K
Device Lot NumberNG3282937H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2023
Initial Date Manufacturer Received 09/20/2023
Initial Date FDA Received10/16/2023
Supplement Dates Manufacturer Received11/06/2023
Supplement Dates FDA Received12/01/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/29/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-