Model Number N/A
Device Problem Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problems Cyst(s) (1800); Bone Fracture(s) (1870); Pain (1994)
Event Date 09/21/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).D10: oxf twin-peg cmntd fem lg pma; ref#161470; lot#056160.Oxf anat brg lt lg size 5 pma; ref#159556; lot#2944765.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported that approximately 9 years and 5 months post-implantation, the patient underwent revision surgery from a partial knee to a total knee due to pain, aseptic loosening, tibial bone collapse, and a cyst forming.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations.The device is used for treatment.Radiographs were provided and reviewed by a radiologist.There is a medial unicompartmental arthroplasty with a medially subsided and loose tibial implant.Radiolucency is present at the central tibial plateau.There is mild knee varus.Loosening and subsidence of the tibial implant of the medial unicompartmental arthroplasty.With the available information, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF UNI TIB TRAY SZ D LM PMA
Type of Device
KNEE PROSTHESIS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key17938000
MDR Text Key325676791
Report Number3002806535-2023-00350
Device Sequence Number1
Product Code NRA
UDI-Device Identifier05019279388905
UDI-Public(01)05019279388905(17)240203(10)945640
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/03/2024
Device Model NumberN/A
Device Catalogue Number154724
Device Lot Number945640
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/21/2023
Initial Date FDA Received10/16/2023
Supplement Dates Manufacturer Received10/25/2023
Supplement Dates FDA Received10/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/03/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexMale
-
-