• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. FAN SPRAY KIT; DISPOSABLE BATTERY-POWERED SURGICAL/ORTHOPAEDIC LAVAGE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. FAN SPRAY KIT; DISPOSABLE BATTERY-POWERED SURGICAL/ORTHOPAEDIC LAVAGE SYSTEM Back to Search Results
Model Number N/A
Device Problems Battery Problem (2885); Activation Problem (4042)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/25/2023
Event Type  malfunction  
Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the unit does not work at all in either high or low mode.Battery was leaking.There is no harm or delay reported.The event timing was during surgery.No adverse events were reported as a result of this malfunction due diligence is complete and no additional information is available.
 
Manufacturer Narrative
G2: foreign country - japan.Functional testing of the returned product found the device powered on, primed in under 10 seconds, and functioned normally in both high and low modes when connected to the original battery pack.No issue or damage was found.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.A definitive root cause cannot be determined as there was no problem found.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
There is no additional information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAN SPRAY KIT
Type of Device
DISPOSABLE BATTERY-POWERED SURGICAL/ORTHOPAEDIC LAVAGE SYSTEM
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key17941378
MDR Text Key325834501
Report Number0001526350-2023-01338
Device Sequence Number1
Product Code FQH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00515047501
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/01/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-