ABBOTT GALLANT DR; NO MATCH Back to Search Results
Model Number CDDRA500Q
Device Problems Over-Sensing (1438); Pacing Problem (1439); Pacing Asynchronously (1441)
Patient Problems Arrhythmia (1721); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2023
Event Type  malfunction  
Manufacturer Narrative
Session records were provided for review by technical services.Analysis of the session records indicated that the pacemaker mediated tachycardia (pmt) and far field r-wave oversensing event was sensed by the device, leading to inappropriate automatic mode switch (ams).
 
Event Description
It was reported that during a follow up in clinic, pacemaker mediated tachycardia (pmt) and far field r-wave oversensing resulting in inappropriate automatic mode switch (ams) was noted on the device.Abbott technical support was contacted and reprogramming of the device was recommended.No intervention has been performed at this time.The patient was in stable condition and will continue to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT DR
Type of Device
NO MATCH
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17943996
MDR Text Key325769488
Report Number2017865-2023-49808
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032003
UDI-Public05415067032003
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDDRA500Q
Device Lot NumberP000172647
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR.; TENDRIL STS.
-
-