ST. JUDE MEDICAL PACEL BIPOLAR PACING CATHETER; ELECTRODE, PACEMAKER, TEMPORARY Back to Search Results
Model Number 401771
Device Problem Material Too Rigid or Stiff (1544)
Patient Problem Cardiac Perforation (2513)
Event Type  Injury  
Event Description
During a procedure for a patient with st-segment elevation myocardial infarction with complete block, a pericardial effusion occurred which required a pericardiocentesis and surgery to stabilize the patient.The pacing catheter was positioned as temporary and a stent was placed on the circumflex coronary artery, when a drop in blood pressure was observed.An echocardiogram was performed and confirmed a pericardial effusion.A pericardiocentesis was done, but bleeding continued because the temporary device moved and created a small hole in the wall.The patient was sent to surgery in parma.The presence of the hole in the ventricular wall was confirmed and the patient stabilized.The physician believes the tip of the catheter was too stiff which may have caused the perforation.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACEL BIPOLAR PACING CATHETER
Type of Device
ELECTRODE, PACEMAKER, TEMPORARY
Manufacturer (Section D)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345 2126
Manufacturer (Section G)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345 2126
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key17944214
MDR Text Key325795223
Report Number2182269-2023-00047
Device Sequence Number1
Product Code LDF
UDI-Device Identifier05414734001137
UDI-Public05414734001137
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K152784
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number401771
Device Lot Number8944979
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/16/2023
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/16/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Life Threatening;
-
-