Catalog Number SL-2000M2095 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/25/2023 |
Event Type
malfunction
|
Event Description
|
As reported by the user facility: detailed inquiry description: it was discovered that there was a blood leak at the junction between the blood pump and the dialyzer at the point where the "pink" capped connecter attaches to the dialyzer.No injury reported.
|
|
Manufacturer Narrative
|
This report has been identified as b.Braun medical internal report number (b)(4).No sample and/or lot number were provided.Further investigation of the complaint is not possible without a sample and/or lot number.The actual defective device is valuable tool in investigating the cause of this incident.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
|
Manufacturer Narrative
|
This report has been identified as b.Braun medical internal report number (b)(4).A review of the manufacturing records for sl-2000m2095 lot # 30155005 was performed and indicated there were no quality issues during the manufacturing of this lot.No sample was provided for evaluation.Based on the data from the investigation we are unable to determine the root cause of the reported incident.The reported defect was unable to be confirmed.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
|