The presence of a ventilator failure was confirmed during on-site check and traced back to the ventilator motor which was worn after almost 11 years of use.In fact, the log file of the device indicates that it has never been switched off after putting it into use for the first time.By doing so, there is a permanent flow of electrical current through the motor; the device adapts the piston position to each change of atmospheric pressure during standby.The motor is designed to last for a minimum of 10 years at standard operation conditions and, post market surveillance data confirms that this is fulfilled in general.It is however not known which long-term effects result from the permanent exposition to current flow.The case is considered a single event.The typical wear effect is that the collector disc becomes worn which leads to positions at the circumference where the electrical contact to the carbon brushes is interrupted and no mechanical power will be provided to the rotor.This leads to speed fluctations and can cause a divergence between the expected piston position and the measured one.To protect the patient from potentially hazardous output the device is designed to shut down automatic ventilation and to alert the user to this condition by means of a corresponding alarm.Manual ventilation via the built-in breathing bag remains possible including gas dosage.
|